FIXED GAS SAMPLING PUMP | RP12 BLDC-R1 | HANLA

VESSEL : FIX GAS SAMPLING PUMP

PORT : KEPPEL BENOI

JOBNO :20140708

EQUIPMENT : FIXED GAS SAMPLING SUCTION PUMP

DATE : 25/07/14

MAKE : HANLA

MODEL : HGS 100

ITEM

DESCRIPTION / REMARKS : CALIBRATION

1

TO CHECK AND CALIBRATE GAS SAMPLING SYSTEM.

 

Health check the system pcb card for analyzing and display unit working in order.

 

Checked the suction pump condition, found suction not satisfactory

 

 

Fix gas detection sampling pump

2

Fix Gas sampling pump suction pump

 

Pump faulty and renewed.

3

fix gas detection display unit

 

Flow calibration.

4

 

hgs-100 display unit

 

Calibrated the HC optima plus sensor by span gas.

5

Gas sampling display unit

 

After calibration, system running properly.