VESSEL : FIX GAS SAMPLING PUMP |
PORT : KEPPEL BENOI |
JOBNO :20140708 |
|
EQUIPMENT : FIXED GAS SAMPLING SUCTION PUMP |
DATE : 25/07/14 |
||
MAKE : HANLA |
MODEL : HGS 100 |
||
ITEM |
DESCRIPTION / REMARKS : CALIBRATION |
||
1 |
TO CHECK AND CALIBRATE GAS SAMPLING SYSTEM.
Health check the system pcb card for analyzing and display unit working in order.
Checked the suction pump condition, found suction not satisfactory
|
||
2 |
Pump faulty and renewed. |
||
3 |
Flow calibration. |
||
4 |
Calibrated the HC optima plus sensor by span gas. |
||
5 |
After calibration, system running properly. |
||