VESSEL : CARGO MONITORING SYSTEM |
PORT : KEPPEL GUL |
JOBNO :20140603 |
|
EQUIPMENT :CARGO MONITORING SYSTEM LMS |
DATE : 17/07/14 |
||
MAKE :HANLA/ENRAF |
MODEL : EM 540 |
||
ITEM |
DESCRIPTION / REMARKS : |
||
1 |
TO CHECK AND CALIBRATE CARGO TANK RADAR LEVEL, TEMPERATURE AND PRESSURE.
HEALTH CHECK THE SYSTEM FOUND ALL OK EXCEPT 1’S LOWER TEMPERATURE DISPLAYED FAIL, LOCAL SIDE SENSOR FAULTY.
|
||
2 |
READ ALL THE CURRENT VALVE FROM RADAR AND COMPARE WITH TANK TABLE FOUND SOME OF THE INPUTTED VALUES ABNORMAL.
MANUALLY ACTIVATE THE RADAR AND READ EMPTY TANK SOUNDING. CALCULATE THE OFFSET BY SOUNDING TABLE AND STORE IT TO RADAR ONE BY ONE. |
||
3 |
CHECKED ALL PRESSURE DISPLAYED ERROR READING, ADJUSTED ALL THE PRESSURE RANGE AND OFFSET.
|
||
4 |
AFTER ADJUSTED ALL THE PARAMETER SETTINGS, LMS DISPLAYING CORRECT READING EXCEPT 1’S TEMPERATURE SENSOR.
|
||
5 |
RECOMMENDATION :
NEED TO RENEW 1’S LOWER TEMPERATURE SENSOR. |
||