HGS-ANNUAL CALIBRATION

    HANLA FIXED GAS SAMPLING SYSTEM - ANNUAL CALIBRATION 

VESSEL : HANLA ANNUAL CALIBRATION HGS-100

PORT : KEPPEL

JOBNO :20140815

EQUIPMENT : FIXED GAS SAMPLING SYSTEM HGS-100

DATE : 01.12.2014

MAKE : HANLA

MODEL : HGS-100

FIXED GAS SAMPLING SYSTEM HGS-100

The Fixed gas sampling system is dedicated to the Gas Control in all Tanks, Void Spaces, Pump room or Houses adjacent to Cargo Storage Tanks and Handling Systems. In order to detect any Gas Concentration level over programmable limits and to monitor visible and audible alarms consequently, these areas are controlled by the suction process sampling on one or more common sensor(s) as well as by individual local sensors.
This system complies with ISGOTT regulation chap. 7.8 and 8.2.

ITEM

DESCRIPTION / REMARKS :

1

JOB DESCRIPTION:

To health check and calibration

FIX GAS SAMPLING DETECTION SYSTEM HGS-100

2

FUNCTION TEST/TRUBLESHOOT :

Durring the healthcheck Found Hc sensor for pump room is not working .Renewed new sensor element and calibrated.

HANLA OXYGEN SENSOR O2

FIXED GAS DETECTION SYSTEM - SERVICE

4

CALIBRATION :

Calibrated the system with span gas.

HANLA SPAN GAS

After done all the settings and calibration,

Tested the system, found working satisfactory.

Copyright © 2016 raycOntrol pte ltd. All rights reserved. All brands and product names are trademark of their respective owners. raycOntrol pte ltd assumes no responsibility or liability for any misinterpretation, loss or damage incurred as a result of any use of the information and material contained within or downloaded from this website. All parties concerned are encouraged to conduct their own information verification or confirmation.