VESSEL : HANLA ANNUAL CALIBRATION HGS-100 |
PORT : KEPPEL |
JOBNO :20140815 |
|
EQUIPMENT : FIXED GAS SAMPLING SYSTEM HGS-100 |
DATE : 01.12.2014 |
||
MAKE : HANLA |
MODEL : HGS-100 |
||
FIXED GAS SAMPLING SYSTEM HGS-100
The Fixed gas sampling system is dedicated to the Gas Control in all
Tanks, Void Spaces, Pump room or Houses adjacent to Cargo Storage Tanks and
Handling Systems. In order to detect any Gas Concentration level over
programmable limits and to monitor visible and audible alarms consequently,
these areas are controlled by the suction process sampling on one or more common
sensor(s) as well as by individual local sensors. |
|||
ITEM |
DESCRIPTION / REMARKS : |
||
1 |
JOB DESCRIPTION:
To health check and calibration
|
||
2 |
FUNCTION TEST/TRUBLESHOOT :
Durring the healthcheck Found Hc sensor for pump room is not working .Renewed new sensor element and calibrated.
|
||
4 |
CALIBRATION :
Calibrated the system with span gas.
After done all the settings and calibration,
Tested the system, found working satisfactory.
|
||
Copyright © 2016 raycOntrol pte ltd. All rights reserved. All brands and product names are trademark of their respective owners. raycOntrol pte ltd assumes no responsibility or liability for any misinterpretation, loss or damage incurred as a result of any use of the information and material contained within or downloaded from this website. All parties concerned are encouraged to conduct their own information verification or confirmation. |
|||